Line 482: | Line 482: | ||
<img src="https://static.igem.org/mediawiki/igem.org/e/ef/T--Technion_Israel-HardwarePDMSchip.jpg" class="img-responsive img-center" width="450" style="cursor: pointer;"> | <img src="https://static.igem.org/mediawiki/igem.org/e/ef/T--Technion_Israel-HardwarePDMSchip.jpg" class="img-responsive img-center" width="450" style="cursor: pointer;"> | ||
</a> | </a> | ||
− | <p class="text-center"><b>Fig. | + | <p class="text-center"><b>Fig. 8:</b> PDMS chip prototype.</p> |
</div> | </div> | ||
</div> | </div> | ||
Line 496: | Line 496: | ||
<img src="https://static.igem.org/mediawiki/igem.org/a/a5/T--Technion_Israel-HardwareFFchip.jpg" class="img-responsive img-center" width="450" style="cursor: pointer;"> | <img src="https://static.igem.org/mediawiki/igem.org/a/a5/T--Technion_Israel-HardwareFFchip.jpg" class="img-responsive img-center" width="450" style="cursor: pointer;"> | ||
</a> | </a> | ||
− | <p class="text-center"><b>Fig. | + | <p class="text-center"><b>Fig. 9:</b> Fluidic Factory chip.</p> |
</div> | </div> | ||
</div> | </div> | ||
Line 548: | Line 548: | ||
<img src="https://static.igem.org/mediawiki/igem.org/0/01/File-T--Technion_Israel-Hardwarefigure8.jpg" class="img-responsive img-center" style="cursor: pointer;"> | <img src="https://static.igem.org/mediawiki/igem.org/0/01/File-T--Technion_Israel-Hardwarefigure8.jpg" class="img-responsive img-center" style="cursor: pointer;"> | ||
</a> | </a> | ||
− | <p class="text-center"><b>Fig. | + | <p class="text-center"><b>Fig. 10:</b> Schematic diagram of the quantitative system </p> |
</div> | </div> | ||
</div> | </div> | ||
Line 558: | Line 558: | ||
<img src="https://static.igem.org/mediawiki/igem.org/f/f8/T--Technion_Israel-HardwarespecsQS.JPG" class="img-responsive img-center" style="cursor: pointer;"> | <img src="https://static.igem.org/mediawiki/igem.org/f/f8/T--Technion_Israel-HardwarespecsQS.JPG" class="img-responsive img-center" style="cursor: pointer;"> | ||
</a> | </a> | ||
− | <p class="text-center"><b>Fig. | + | <p class="text-center"><b>Fig. 11:</b>3D Model of the quantitative system </p> |
</div> | </div> | ||
</div> | </div> | ||
Line 571: | Line 571: | ||
<br> | <br> | ||
<br> | <br> | ||
− | To avoid undesired light reflections, we have designed a dedicated black box as shown in Fig | + | To avoid undesired light reflections, we have designed a dedicated black box as shown in Fig 11, to house the chip and the electrical circuits discussed below. |
<br> | <br> | ||
Line 595: | Line 595: | ||
<div class="col-md-6 col-sm-12 vcenter"><!--6 text--> | <div class="col-md-6 col-sm-12 vcenter"><!--6 text--> | ||
<p class="text-justify"> | <p class="text-justify"> | ||
− | The system consists of two independent electrical circuits as shown in Fig | + | The system consists of two independent electrical circuits as shown in Fig 12. The blue circuit contains a resistor of 10 [kΩ], a potentiometer, a LED and Arduino as a constant voltage source. The red circuit contains a photoresistor LDR that is sensitive to 600 [nm] wavelength, a resistor of 1[MΩ] and an Arduino controller. The Arduino supplies constant voltage to both circuits and measures the voltage that falls on the 1[MΩ] resistor. |
</p> | </p> | ||
Line 604: | Line 604: | ||
<img src="https://static.igem.org/mediawiki/igem.org/c/c0/File-T--Technion_Israel-Hardwarefigure10.jpg" class="img-responsive img-center" style="cursor: pointer;"> | <img src="https://static.igem.org/mediawiki/igem.org/c/c0/File-T--Technion_Israel-Hardwarefigure10.jpg" class="img-responsive img-center" style="cursor: pointer;"> | ||
</a> | </a> | ||
− | <p class="text-center"><b>Fig. | + | <p class="text-center"><b>Fig. 12:</b> Schematic diagram of the two electrical circuits.</p> |
</div> | </div> | ||
</div> | </div> | ||
Line 622: | Line 622: | ||
<div class="col-md-6 col-sm-12 vcenter"><!--6 text--> | <div class="col-md-6 col-sm-12 vcenter"><!--6 text--> | ||
<p class="text-justify"> | <p class="text-justify"> | ||
− | The Arduino controller collects samples of the voltage that falls on the 1[MΩ] resistor. The voltage is converted to a digital signal and fed to the computer. The computer data system is based on a <a href="https://static.igem.org/mediawiki/igem.org/b/b1/File-T--Technion_Israel-HardwareGUI.zip">"Matlab GUI" </a>. Note that the Arduino I/O toolbox needs to be installed. <br>When running the Matlab code, the window shown in Fig | + | The Arduino controller collects samples of the voltage that falls on the 1[MΩ] resistor. The voltage is converted to a digital signal and fed to the computer. The computer data system is based on a <a href="https://static.igem.org/mediawiki/igem.org/b/b1/File-T--Technion_Israel-HardwareGUI.zip">"Matlab GUI" </a>. Note that the Arduino I/O toolbox needs to be installed. <br>When running the Matlab code, the window shown in Fig 13 pops up. |
</p> | </p> | ||
Line 630: | Line 630: | ||
<img src="https://static.igem.org/mediawiki/igem.org/0/0d/File-T--Technion_Israel-Hardwarefigure11.jpg" class="img-responsive img-center" style="cursor: pointer;"> | <img src="https://static.igem.org/mediawiki/igem.org/0/0d/File-T--Technion_Israel-Hardwarefigure11.jpg" class="img-responsive img-center" style="cursor: pointer;"> | ||
</a> | </a> | ||
− | <p class="text-center"><b>Fig. | + | <p class="text-center"><b>Fig. 13:</b> The system's user interface</p> |
</div> | </div> | ||
</div> | </div> | ||
Line 811: | Line 811: | ||
</div> | </div> | ||
</div> | </div> | ||
+ | <br> | ||
<!-- 12 text div --> | <!-- 12 text div --> | ||
Line 882: | Line 883: | ||
<img src="https://static.igem.org/mediawiki/igem.org/2/2c/File-T--Technion_Israel-Hardwarefigure12.png" class="img-responsive img-center" style="cursor: pointer;"> | <img src="https://static.igem.org/mediawiki/igem.org/2/2c/File-T--Technion_Israel-Hardwarefigure12.png" class="img-responsive img-center" style="cursor: pointer;"> | ||
</a> | </a> | ||
− | <p class="text-center"><b>Fig | + | <p class="text-center"><b>Fig 14</b> The output voltage for different values of O.D as a function of time. |
<br> | <br> | ||
<br> | <br> |
Revision as of 11:24, 19 October 2016
Hardware
Introduction
We developed our first FlashLab prototypes around on the commercial fluidic chip -IBIDI sticky–Slide I Luer 0.8. This chip is designed mostly for performing cell culture experiments with a custom specific bottom.
The commercial chip worked for preliminary testing but was not idle for our uses: The entry slots are relatively wide, making it difficult to load the sample in a uniform and even fashion.
This affects the diffusion of the chemo-repellent in the channel and reduces the overall accuracy of the device. Also, the channel is relatively shallow, forcing the use of a high concentration of bacteria to get a visible signal, this proves to be a problem as storing a large amount of bacteria in a confined space might cause oxygen shortage that will harm bacterial motility.
.
We devised new solutions to confronts those problems:
- Design a novel chip, based on the commercial chip but with unique changes to its geometry for improved performance.
- Design a quantitative test, a device built especially for detecting a change in bacterial concentration in the chip. This device is much more sensitive than the naked eye.
Chip redesigned
We designed a new chip with the following improvements:
(a) Reducing the radius of the entry slot will enable a controlled insertion of the sample.
The smaller slot will slow down any flow (for example, flow caused by loading a sample from a syringe). Also, this will fix the diffusion source at a permanent place for all of our experiments.
(b)Shaping the channel as a funnel in order to concentrate the bacteria even further as they move away from chemo-repellents (from left to right).
(c) A deeper channel will result in a darker shade of color in the same bacterial concentration than in the commercial chip, while reducing the risk of oxygen shortage.
The new chip was fabricated in two methods: as a PDMS chip and in a Dolomite 3D printer.
Fabricating the PDMS chip
PDMS is considered the standard for microfluidic fabrication in labs. It is optically clear, and in general, inert, non-toxic, and non-flammable.
The PDMS was then fabricated according to the following steps:
1. Design a two part mold using SolidWorks software- cover and base.
2. Print the mold using Ultimaker 2 Etentended+ 3D printer.
3. Mix the polymer base and curing agent at 10:1 weight ratio, respectively. Then, fill the mold with the mix.
4. Place the mold inside a desiccator to degas for 2 hours.
5. Bake the mold at 70 C for 3 hours.
6. Carefully take off the mold’s cover and then cut out the PDMS chip.
7. Attach the PDMS chip to a thin cover glass (0.3 mm) using silicon glue*.
The following scheme describes the mentioned process:
*Traditionally, bonding PDMS to glass is done by plasma treatment. Our 3D printed mold resulted in PDMS chips with relatively rough surface finish, forcing us to use other methods.
Designing the mold
The mold is comprised of two parts which together create a unique geometry and allow for
easier extraction of the PDMS out of the mold.
The base
- The cone on the base of the floor is make the funnel shape of the chip ((a) in figure 3).
- Small slits were made in the walls of the base to position the cover accurately.
- The overall size was determent so the chip will fit on standard microscope cover
slide. this will enable us to run experiments under a microscope easily.
The cover
- Four rods coming out of the sides of the cover for easy extraction of the cover
when taking out the PDMS.
- The ramp is to insure that the channel will be inserted inside the PDMS and
getting the wanted channel height.
- The cover is made smaller than the base for a good fit and for letting out
any gas that might have been caught when inserting it. Those gases, if
left in will expend in the oven and cause deformation in the chip.
Printing the mold using Ultimaker 2 Etentended+. This 3d printer was chosen because of its high accuracy (X,Y,Z =12.5, 12.5, 5 micron) and due that the fact that the polymer it uses (PLA) can be heated to relatively high temperatures without changing form (TG=60-65 C) and does not reacts to the PDMS. Another benefits of 3D printing are the low price and fast manufacturing time: We printed our mold for about 25$, and it took about 6 hours.
Dolomite Fluidic Factory
Factory enables fast prototyping of microfluidic chips, manifolds and connectors using COC (FDA approved, biocompatible, translucent and robust polymer). Printing the chip toke about 3 hours and was made straight from computer model. This technology just came out this year and we are the first iGEM group to ever use it.
Results
We were able to make a few prototypes of the PDMS chip. The extraction of the chip was relatively easy and without any visible cracks or deformations. The chip still needed to be punctured in the entry slots, duo to spaces between the molds. Also, attaching a glass slide to the PDMS needed to be done carefully, as the glass is thin and brittle.
Fig. 8: PDMS chip prototype.
Quantitative test for bacterial concentration
Principle of Operation
Our system uses a photo sensor to measure the intensity of a light beam transmitted through the chip. The measurement process is as follows: A yellow LED emits light at 585-595 [nm] on the chip, with the bacteria inside absorbing a portion of the light. The light transmitted through the chip reaches the photo sensor which outputs an analog signal. This signal is then translated to a digital signal and fed to the computer. The end result is a graph of the output voltage as a function of time.
The output voltage can be compared to the bacterial concentration as shown in section 3.3.
The system requires two measurements. The first measurement is a blank meant to calibrate the system. This measurement is done on a chip containing only motility buffer (control). The second measurement is for the bacterial solution.
To avoid undesired light reflections, we have designed a dedicated black box as shown in Fig 11, to house the chip and the electrical circuits discussed below.
The electrical circuits
The system consists of two independent electrical circuits as shown in Fig 12. The blue circuit contains a resistor of 10 [kΩ], a potentiometer, a LED and Arduino as a constant voltage source. The red circuit contains a photoresistor LDR that is sensitive to 600 [nm] wavelength, a resistor of 1[MΩ] and an Arduino controller. The Arduino supplies constant voltage to both circuits and measures the voltage that falls on the 1[MΩ] resistor.
Computer data system
The Arduino controller collects samples of the voltage that falls on the 1[MΩ] resistor. The voltage is converted to a digital signal and fed to the computer. The computer data system is based on a "Matlab GUI" . Note that the Arduino I/O toolbox needs to be installed.
When running the Matlab code, the window shown in Fig 13 pops up.
The relation between the resistor’s voltage and bacterial O.D.
According to the voltage divider rule, the voltage that falls on the 1[MΩ] resistor VR is equal to
According to “Emant”, the relationship between the resistance RL of a typical LDR and the light intensity is:
Where Iis the light intensity that reaches the photoresistor
Combining equations 1 and 2 we receive:
By definition:
Where A is the optical density of the sample and I0 is the light intensity emitted from the LED
Rearranging Equation 4
Integrating Equation 5 in Equation 1:
Where I0 is the light intensity emitted from the LED and A is the optical density of the bacterial concentration inside the chip.
From Equation 6 it can be derived that VR is expected to decrease as A increases.
System improvements
Initially the photo sensor we intended to use was a photodiode. Since the photodiode is relatively big it was difficult to fix its position. Thus, we replaced it with a photo resistor which is smaller and relatively sensitive to 600 [nm] wavelength.
In addition, before building the final system we used a battery as a voltage source and a USB data acquisition from NI to convert the analog signal into a digital one. In order to improve the system we replaced those two components with an Arduino controller that can serve as a constant voltage source and as a converter simultaneously.
As mentioned before the chip and the two electrical circuits were placed in a dark box (As shown in Fig XX) to avoid undesired light scattering. All the sides of the chip were darkened as well so the light can be transmitted only through the transparent channel.
To improve the dynamic range of the photoresistor we connected a resistor in series with the photo resistor. When the photoresistor is exposed to high light intensity, its resistance decreases dramatically. Under these conditions, most of the voltage falls on the resistor. Since
VR increases with R. As we wanted the maximum voltage falling on the resistor to be 5v (the total voltage), we chose a resistor of 1[MΩ].
Finally, if the light intensity that originates from the LED is too high, it can lead to the saturation of the photo resistor. To be able to tune the light intensity of the LED, a potentiometer was added to the LED circuit, to adjust the desired resistance which produces the optimal light intensity.
Testing the system
As can be deduced from the mathematical equations the voltage is expected to decrease as the optical density increases. For that purpose we prepared bacterial solutions in motility buffer at different concentrations and loaded them to the system. the results are displayed in Fig XX.
As can be deduced from the graph the output voltage converges after 88 [sec] which is much less than the time required for cluster formation (about 15 minutes). Thus, the system indeed can be used with FlashLab for real time detection. Moreover, the dynamic range of the system is relatively wide (0-3v), giving us the ability to detect a variety of bacterial O.D levels. In addition, the difference between the outputs obtained for O.D 0.757 and O.D 0.653 is much bigger than the system error’s measurement. Hence, it can be concluded that the system is relatively precise.
Overview
FlashLab, although a successful detection tool, has several drawbacks. By redesigning the fluidic channels and engineering a more sensitive measurement system, we will be able to get a more reliable, accurate and user friendly device.
Our preliminary testing supports those claims. Showing we can detect small differences in bacterial concentration, that is not noticeable otherwise. Our prototype was completed by designing an easy to use user interface, and making a more reliable and cost effective system. We believe this device can have a real world, commercial potential.
In the future,we plan first to expand our testing and improve the chip even more. Ideally, to design a chip that is compatible with different tests (for fast/slow moving, high concentrations of repellent, different temperatures etc.). Second, we plant to improve the quantitative device, by replacing to a more accurate sensor or adding by implementing a signal processing algorithm for better results. Third, according to the model we developed, there is a clear correlation between the repellent gradient and the bacterial concentration. The system allows to get a quantitative estimation of the bacterial concentration, so theoretically the results can be correlated to the repellent/attractant concentration.